Operating Instructions for the ELIT 003n
Lithium Acetate (CH3COOLi)
Double Junction Reference Electrode
Introduction
The ELIT 8mm reference electrode is similar in principle to a conventional
double junction reference electrode in that it has two chambers and two junctions. But,
unlike glass electrodes, the filling solutions are in the form of gel rather than liquid.
Thus there is very little loss of electrolyte during use and the plastic body is a sealed unit
which is very robust and long-lasting and does not require any maintenance.
Installation
This electrode is designed specifically for use in combination with an ELIT 8mm ion
selective electrode and must be connected to the measuring system by inserting into a Dual
Electrode Head (ELIT 201) together with a suitable ISE. This system replaces
conventional Combination Electrodes.
Storage and Operation
All reference electrodes should be kept immersed in a solution with the same composition
as the outer filling solution when not in use. Thus they are shipped and stored with the
ceramic frits protected by a plastic cap containing this solution - in this case, 0.1 Molar CH3COOLi (but note that this concentration is not critical, say ± 10%). This cap must be carefully
removed so as not to spill the solution and kept upright for replacing when the electrode is
not in use. The electrode tip must then be rinsed with a jet of de-ionised water and gently
dabbed dry with a lint-free laboratory tissue before use.
After use the protective cap must be topped up with solution if necessary and replaced to prevent the ceramic frit from drying out and becoming blocked.
NOTE: It is strongly recommended that the electrode is removed from the electrode head
before attempting to remove or replace the cap, in order to prevent unnecessary force being
applied to the joint between the electrode body and the connecting pin. For more information about reference electrodes and Ion Selective Electrode measurements, please visit our website: www.nico2000.net.
|