Return to Ion Selective Electrodes page
NOTE: Many published ISE methods for ammonium use an ammonia gas-sensing electrode to measure the NH3 gas liberated after the sample has been treated with strong alkali. Details of analytical method Links to relevant publications on the WWW
Technical Specifications Introduction Physical Specifications Chemical / Operational
Specifications The following ions interfere with the ammonium
measurement - Selectivity Coefficients (SC) in brackets: Potassium (0.11), Sodium
(0.0016), Magnesium (0.00018), Calcium (0.00006), Lithium (0.00003). The SC is the approximate apparent increase in the measured
concentration caused by 1 unit of the interferent. Thus the likely effect of any
interfering ion (% increase) can be calculated as follows: ((expected concentration) x (SC) /
(expected NH4 concentration)) x 100. On this basis, only the K interference is likely to
cause a significant error. Return to top
Note 1: Because of the interference of K+ on the NH4+ electrode, special techniques are required for measuring samples with a K content greater than half that of the NH4 - see Procedure 2 Note 2: In aqueous solution, there is an equilibrium between NH4+ and NH3 which is pH dependent. For more information, see: Apparatus Required: Ion-Selective Electrode for ammonium ion (ELIT 8051 PVC membrane). Reference electrode: double junction lithium acetate (ELIT 003). Dual electrode head (ELIT 201). ELIT Computer Interface/Ion Analyser, or Ion/pH/mV meter. Standard solution: 1000 ppm NH4 as NH4Cl . Buffer solution (ISAB): 1 Molar CuSO4. 100 or 150 ml polypropylene beakers, 100ml volumetric flask, 1, 2, 5, 10ml pipettes. Standard Solution Preparation Calibration: Before use, the electrodes must be calibrated by measuring a series of known standard solutions, made by serial dilution of the 1000ppm standard solution. For a full calibration, prepare 100ml of solutions containing 1000, 100, 10, 1, and 0.1ppm NH4. If the approximate range of concentrations of the samples is known, and this is within the specified linear range of the ISE, then it is only necessary to make two solutions which span this range: e.g. if samples are known to lie between, say, 30 and 130ppm then you could use standards of 10 and 200ppm or even 20 and 150ppm. Follow the instructions on the ELIT Ion Analyser software to measure these standard solutions and prepare a calibration graph. NB: If the samples to be measured are expected to have a total ionic strength
of greater than 0.01 Molar, then add 10 ml of buffer solution to each standard
and mix thoroughly to compensate for different activity coefficients between
samples and standards. BUT note that this will change the calibration slightly (ie: increase the slope but also raise the lower limit of linearity and detection limits) and lower the pH so that any NH3 will be converted to NH4. Sample Preparation: For low ionic strength samples, no sample preparation is necessary. Simply take approximately 50 to 100 mls of sample in a plastic beaker, or even immerse the electrodes directly in a lake or river (but take care to avoid losing the electrodes!). For high ionic strength samples, take 100mls of sample and add 10 mls of buffer solution and stir well before measurement. Sample Measurement: Follow the instructions in the computer interface software and/or electrode operating instructions to measure a series of samples and record the results. Briefly, it is important to note that, if measuring in beakers, the electrodes must be washed and dried between each sample, to avoid cross contamination, and sufficient time must be allowed (2 or 3 minutes), before taking a reading after immersion, to permit the electrode signal to reach a stable value. For the highest precision, frequent recalibration is recommended (see operating instructions). Results: The results will be displayed as ppm and mol/l. If buffer solution has been added equally to standards and samples then these figures will not need adjusting because they will all be affected by the same dilution factor.
Return to top Procedure for determining the concentration of AMMONIUM
(NH4+) in aqueous solutions with high potassium
content. For samples with K/NH4 ratios greater than about 0.5, rather than relying on the general Selectivity Coefficient quoted in the electrode specifications to assess the likely level of interference, a more reliable correction can be made by measuring the interference directly in a typical sample. i.e. first measure the NH4 and K in the sample, then add more K (sufficient to ensure a significant increase in signal - eg double the concentration) and measure apparent NH4 again. Then calculate the SC from the increase in NH4 caused by known addition of K. However, it is recommended that the accuracy and precision of this correction should be evaluated for each particular application before too much reliance is placed on it. This measurement can be made, with or without ISAB, by Direct Potentiometry or Standard Addition depending on the Ionic Strength of the samples. Alternatively, a more accurate correction can be made using the following method which involves measuring the ammonium and potassium simultaneously with a triple electrode head and multi-channel ion analyser, and making appropriate corrections using neural network software to interrogate an extensive built-in calibration data-base. It is currently only suitable for samples with a total ionic strength of < 0.01 Molar and with ammonium contents in the range 0.1 to 100 ppm and potassium up to 100ppm. NOTE: this apparatus and software is a prototype which is not yet commercially available. If you are interested in this product and prepared to undertake some collaborative on-site testing to identify any un-detected problems for your particular application, please e-mail info@nico2000.net OR nico2000ltd@btinternet.com for further information.For a detailed discussion, please
see: Apparatus Required: Ion-Selective Electrode for ammonium ion (ELIT 8051 PVC membrane) Potassium ISE for monitoring ionic interference (ELIT 8031 - PVC) Reference Electrode: double junction lithium acetate (ELIT 003). Triple electrode head to hold two ISEs and one RE: (ELIT 301). Two channel interface (ELIT 8902) and desk-top or lap-top computer. ELIT Special Ammonium Analyser Software. Standard ammonium solution: 1000 ppm NH4 as NH4Cl. Standard potassium solution: 1000 ppm K as KCl. 100 or 150 ml polypropylene beakers, 100ml volumetric flask, 1, 2, 5, 10ml pipettes. Calibration: The system simply requires a single-point recalibration each time it is used, in order to compensate for any day-to-day variations in electrode response due to aging or temperature differences, and any difference between the analysing electrodes and those used for the original calibration. The standard solution must contain known concentrations of both ions and must be identical to one of the standards included in the original data base. For the most precise work, the standard solution should be chosen to be close to the range of concentrations expected for the samples and should be freshly made from concentrated solutions in order to avoid deterioration during storage. Therefore it is recommended that this standard is made up by the operator each day before commencing analysis. Sample Measurement: Normally no sample preparation is necessary. Simply take a sample of between 50 and 100 mls in a plastic beaker and immerse the electrodes (or even immerse directly in a lake or river – but care must be taken to avoid losing the electrodes!) and wait for a stable signal. If using beakers, the electrodes should be washed and dried between each sample to avoid cross contamination. Results: The results are displayed and recorded as ppm of NH4 and K. If samples are found to lie above the concentration range of the instrument then appropriate dilutions can be made with de-ionised water to bring them into range. In this case the results must be adjusted manually to allow for this dilution factor. Return to top |