Return to Ion Selective Electrodes page
|
Method for determining the concentration of MERCURY (Hg++-) in WATER Links to relevant publications on the WWW Technical Specifications Introduction Physical Specifications Chemical / Operational
Specifications Interference: Procedure for determining the concentration of MERCURY (Hg++-) in WATER Apparatus Required: Ion-Selective Electrode for mercuric ion (ELIT 8251 crystal membrane). Reference electrode: double junction potassium nitrate (ELIT 002). Dual electrode head (ELIT 201) Standard solution: 1000 ppm Hg as Hg(NO3)2 Buffer solution (ISAB): 0.1 Molar HNO3. ELIT Computer Interface/Ion Analyser, or Ion/pH/mV meter. 100 or 150 ml polypropylene beakers, 100ml volumetric flask, 1, 2, 5, 10, 25ml pipettes. Calibration: Before use, the electrodes must be calibrated by measuring a series of known standard solutions, made by serial dilution of the 1000ppm standard solution. For a full calibration, prepare 100ml of solutions containing 1000, 100, 10, 1, and 0.1ppm Hg. If the approximate range of concentrations of the samples is known, and this is within the specified linear range of the ISE, then it is only necessary to make two solutions which span this range: e.g. if samples are known to lie between, say, 30 and 130ppm then you could use standards of 10 and 200ppm or even 20 and 150ppm. NB: This electrode can only be used in the pH range 0 to 2 therefore it is essential to add ISAB 50% by volume to all standards and samples before measurement. Follow the instructions on the computer interface software to measure these standard solutions and prepare a calibration graph. Sample Preparation: Take 25mls of sample and add 25 mls of buffer solution and stir well before measurement. Sample Measurement: Follow the instructions in the computer interface software and/or electrode operating instructions to measure a series of samples and record the results. Briefly, it is important to note that the electrodes must be washed and dried between each sample, to avoid cross contamination, and sufficient time must be allowed (2 or 3 minutes), before taking a reading after immersion, to permit the electrode signal to reach a stable value. For the highest precision, frequent recalibration is recommended (see operating instructions). Results: The results will be displayed as ppm and mol/l. If buffer solution has been added equally to standards and samples then these figures will not need adjusting because they will all be affected by the same dilution factor. Return to top |