Return to Ion Selective Electrodes page
|
Method for determining the concentration of THIOCYANATE (SCN-) in WATER Links to relevant publications on the WWW Technical Specifications Introduction Physical Specifications Chemical Specifications Interference: Return to top Procedure for determining the concentration of THIOCYANATE (SCN-) in WATER Apparatus Required: Ion-Selective Electrode for thiocyanate ion (ELIT 8229 crystal membrane) Reference electrode: double junction potassium nitrate (ELIT 002). Dual electrode head (ELIT 201) Standard solution: 1000 ppm SCN- as KSCN Buffer solution (ISAB): 5 Molar NaNO3 . ELIT Computer Interface/Ion Analyser, or Ion/pH/mV meter. 100 or 150 ml polypropylene beakers, 100ml volumetric flask, 1, 2, 5, 10ml pipettes. Calibration: Before use, the electrodes must be calibrated by measuring a series of known standard solutions, made by serial dilution of the 1000ppm standard solution. For a full calibration, prepare 100ml of solutions containing 1000, 100, 10, and 1ppm SCN. If the approximate range of concentrations of the samples is known, and this is within the specified linear range of the ISE, then it is only necessary to make two solutions which span this range: e.g. if samples are known to lie between, say, 30 and 130ppm then you could use standards of 10 and 200ppm or even 20 and 150ppm. NB: If the samples to be measured are expected to have a total ionic strength of greater than 0.01 Molar, then 2 ml of buffer solution should be added to each 100ml standard and mixed thoroughly to compensate for different activity coefficients between samples and standards. Follow the instructions on the computer interface software to measure these standard solutions and prepare a calibration graph. Sample Preparation: For low ionic strength samples, no sample preparation is necessary. Simply take approximately 50 to 100 mls of sample in a plastic beaker, or even immerse the electrodes directly in a lake or river (but take care to avoid losing the electrodes!). For samples with high ionic strength, take 100mls of sample and add 2 mls of buffer solution and stir well before measurement. Sample Measurement: Follow the instructions in the computer interface software and/or electrode operating instructions to measure a series of samples and record the results. Briefly, it is important to note that, if measuring in beakers, the electrodes must be washed and dried between each sample, to avoid cross contamination, and sufficient time must be allowed (2 or 3 minutes), before taking a reading after immersion, to permit the electrode signal to reach a stable value. For the highest precision, frequent recalibration is recommended (see operating instructions). Results: The results will be displayed as ppm and mol/l. If buffer solution has been added equally to standards and samples then these figures will not need adjusting because they will all be affected by the same dilution factor. <Return to top |