Return to  Ion Selective Electrodes page

Click on the following links for details of analytical procedures for determining the concentration of CALCIUM (Ca++) in various substrates:

| WATER | ANIMAL FEED | BEER | FRUIT JUICE | MEAT | SKIMMED MILK | SOIL |

Links to relevant publications on the WWW



Technical Specifications
for the Calcium Ion-Selective Electrode
(ELIT 8041)

Click here to download a printer-friendly MS Word 97 Specification Sheet.

Introduction
The Calcium Ion-Selective Electrode has a solid-state PVC polymer matrix membrane. The electrode is designed for the detection of calcium ions ( Ca+2 ) in aqueous solutions and is suitable for use in both field and laboratory applications.
The Calcium Ion is a divalent cation .
One mole of ( Ca+2 ) has a mass of 40.080 grams; 1000 ppm is 0.025 M

Physical Specifications
Length of body excluding gold contact = 130 mm
Length of body including gold contact = 140 mm
Diameter of body = 8 mm
DC resistance at 25°C < 2.5 MOhm
Minimum feasible sample volume = 5 ml

Chemical / Operational Specifications
Preconditioning / Standard solution : Normally 1000 ppm Ca+2 as CaCl2
(But see General Operating Instructions)
Preconditioning time : at least 5 minutes
Optimum pH range : pH 3.5 to pH 11
Temperature range : 5°C to 50°C
Recommended ISAB : 4M KCL (Add 2% v/v)
Recommended reference electrode : Single junction (ELIT 001)
Electrode slope at 25°C : 26± 3 mV/decade
Concentration range : 0.02 to 40,000 ppm (5x10-7 to 1 Molar)
Response time : < 10 seconds
Defined as time to complete 90% of the change in potential after immersion in the new solution.
Time for stable reading after immersion : < 1 to > 5 minutes
Depending on concentration, use of ISAB, nature of sample and stabilisation time of liquid junction potential of reference electrode.
Potential drift (in 1000 ppm unstirred) < 3 mV/day (8 hours)
Measured at constant temperature and with ISE and Reference Electrode continually immersed

Interference
Only Al +3 has a strong interference on the Ca electrode, with a selectivity coefficient (SC) of 5 (ie five times more sensitive to Al +3 than to Ca +2) and can only be tolerated in very low concentrations relative to the Ca.
The SC is the approximate apparent increase in the measured concentration caused by 1 unit of the interferent.  Thus the likely effect of any interfering ion (% increase) can be calculated as follows: 
     
((expected concentration) x (SC) / (expected Ca concentration)) x 100.

The following ions have a small effect on the Ca electrode (SC in brackets), but these would only be significant if they were at least ten times more concentrated than the Ca ion:  Iron Fe+2 (0.02) Strontium (0.0075), Barium (0.005) Copper (0.002), Sodium, (0.0005), Magnesium (0.00055), Potassium (0.00005), Ammonium (0.000025), Lithium (0.000012)




RETURN TO TOP

Procedure for determining the concentration of CALCIUM (Ca++) in WATER

Apparatus Required:

Ion-Selective Electrode for calcium ion (ELIT 8041 PVC membrane)

Reference electrode: single junction silver chloride (ELIT 001)

Dual electrode head (ELIT 201)

Standard solution: 1000 ppm Ca as CaCl2

Buffer solution (ISAB): 4 Molar KCl.

ELIT Computer Interface/Ion Analyser, or Ion/pH/mV meter.

100 or 150 ml polypropylene beakers, 100ml volumetric flask, 1, 2, 5, 10ml pipettes.

Calibration:

Before use, the electrodes must be calibrated by measuring a series of known standard solutions, made by serial dilution of the 1000ppm standard solution. For a full calibration, prepare 100ml of solutions containing 1000, 100, 10, 1, and 0.1ppm Ca. If the approximate range of concentrations of the samples is known, and this is within the specified linear range of the ISE, then it is only necessary to make two solutions which span this range: e.g. if samples are known to lie between, say, 30 and 130ppm then you could use standards of 10 and 200ppm or even 20 and 150ppm.

NB: If the samples to be measured are expected to have a total ionic strength of greater than 0.001 Molar, then 2 ml of buffer solution should be added to each 100ml standard and mixed thoroughly to compensate for different activity coefficients between samples and standards.

Follow the instructions on the computer interface software to measure these standard solutions and prepare a calibration graph.

Sample Preparation

For low ionic strength samples, no sample preparation is necessary. Simply take approximately 50 to 100 mls of sample in a plastic beaker, or even immerse the electrodes directly in a lake or river (but take care to avoid losing the electrodes!). For samples with high ionic strength, take 100mls of sample and add 2 mls of buffer solution and stir well before measurement.

Sample Measurement

Follow the instructions in the computer interface software and/or electrode operating instructions to measure a series of samples and record the results. Briefly, it is important to note that, if measuring in beakers, the electrodes must be washed and dried between each sample, to avoid cross contamination, and sufficient time must be allowed (2 or 3 minutes), before taking a reading after immersion, to permit the electrode signal to reach a stable value. For the highest precision, frequent recalibration is recommended (see operating instructions).

Results

The results will be displayed as ppm and mol/l. If buffer solution has been added equally to standards and samples then these figures will not need adjusting because they will all be affected by the same dilution factor.





RETURN TO TOP

Procedure for determining the concentration of CALCIUM (Ca++) in ANIMAL FEED

Calcium is extracted from animal feedstuffs by ashing a known weight of sample and dissolving the ash in mixed hydrochloric and nitric acids. The pH is then adjusted to about pH7 with potassium hydroxide.

Apparatus Required

Ion-Selective Electrode for calcium ion (ELIT 8041 – PVC membrane)

Reference electrode: single junction silver chloride (ELIT 001)

Dual electrode head (ELIT 201)

Standard solution: 1000 ppm Ca as CaCl2

Buffer solution (ISAB): 4 Molar KCl.

ELIT Computer Interface/Ion Analyser, or Ion/pH/mV meter.

100 or 150 ml polypropylene beakers, 100ml volumetric flask, 1, 2, 5, 10ml pipettes.

For Sample Preparation:

1Molar HCl, 4 Molar KOH, conc. HNO3

Muffle furnace, balance, platinum or silica crucibles, 250ml volumetric flasks.

pH electrode (ELIT P500)

Calibration

Before use, the electrodes must be calibrated by measuring a series of known standard solutions, made by serial dilution of the 1000ppm standard solution. For a full calibration, prepare 100ml of solutions containing 1000, 100, 10, 1, and 0.1ppm Ca. If the approximate range of concentrations of the samples is known, then the sample weight can be chosen to ensure a solution concentration which is within the specified linear range of the ISE. In this case it is only necessary to make two solutions which span this range: e.g. if sample solutions are known to lie between, say, 30 and 130ppm then you could use standards of 10 and 200ppm.

2 ml of buffer solution must be added to each 100ml standard and mixed thoroughly to compensate for different activity coefficients between samples and standards.

Follow the instructions on the computer interface software to measure these standard solutions and prepare a calibration graph.

Sample Preparation

Weigh out an appropriate quantity of finely ground material to give a solution concentration which lies within the range of the standards chosen. If the approximate sample concentration is completely unknown then try using 1g. Transfer the sample to a suitable crucible and ignite in a muffle furnace at 320°C for 3 hours, then allow to cool. When cool, slowly add 50 ml of 1M HCl, taking care to minimize spitting, and then add 5 drops of conc. HNO3. Transfer this solution to a tall 150ml beaker or conical flask and boil gently for 30 minutes Cool this solution and adjust the pH to about 7 by adding 4M KOH and monitoring the pH change with a pH electrode (previously calibrated) connected to the computer interface. Transfer this solution to a 250ml volumetric flask and dilute to the mark with deionised water. Take 100ml of this solution in a 150ml plastic beaker and add 2 ml of buffer solution and mix well.

Sample Measurement

Follow the instructions in the computer interface software and/or electrode operating instructions to measure a series of samples and record the results. Briefly, it is important to note that the electrodes must be washed and dried between each sample, to avoid cross contamination, and sufficient time must be allowed (2 or 3 minutes), before taking a reading after immersion, to permit the electrode signal to reach a stable value. For the highest precision, frequent recalibration is recommended (see operating instructions).

Results

The results will be displayed as ppm and mol/l in the solution. Since the buffer solution has been added equally to standards and samples then these figures will not need adjusting because they will all be affected by the same dilution factor. To obtain the ppm Ca in the sample (mg/Kg) the ppm in the solution must be multiplied by 250 and divided by the sample weight. For wt% Ca this figure must be divided by 1000.





RETURN TO TOP

Procedure for determining the concentration of CALCIUM (Ca++) in BEER

Because of the complex matrix, high ionic strength, and other interfering factors in beer, the concentration of ions in the solution is best determined using the Standard Addition method rather than direct potentiometry.

Apparatus Required

Ion-Selective Electrode for calcium ion (ELIT 8041 – PVC membrane)

Reference electrode: single junction silver chloride (ELIT 001)

Dual electrode head (ELIT 201)

Standard solution: 1000 ppm Ca as CaCl2

ELIT Computer Interface/Ion Analyser, or Ion/pH/mV meter.

100 or 150 ml polypropylene beakers, 100ml volumetric flask, 1, 2, 5, 10ml pipettes.

Calibration

Before use, the electrodes must be calibrated to find an accurate value for the slope under the current operating conditions and in the range of the expected sample concentrations. If sample concentration is completely unknown then it is best to make a quick sample measurement first, using direct potentiometry with an old calibration. Then make up and measure two standards which span the expected sample range - say 1 & 10ppm or 10 & 100ppm - or more closely spaced if working below the linear range of the electrode.

Standard Selection

Enter the slope and expected sample concentration and volume into the Standard Addition software to calculate an appropriate standard concentration and volume. Make up the calculated standard concentration by serial dilution of the 1000ppm standard.

Sample Measurement

1) Pipette the pre-determined sample volume (usually 50 or 100ml) into a beaker, immerse the electrodes, swirl the solution, and record the stable voltage after a few minutes of stabilization.

2) Add the calculated volume of standard, swirl or stirr well then wait for a new stable voltage when the solutions have completely mixed and the electrodes have re-equilibrated.

The software will then calculate the sample concentration and give an estimate of the quality of the determination - if the sample concentration is very different from the expected value then you will be prompted to make a second measurement using a more appropriate volume or concentration of standard.





RETURN TO TOP

Procedure for determining the concentration of CALCIUM (Ca++) in FRUIT JUICE

Because of the complex matrix, high ionic strength, and other interfering factors in fruit juices, the concentration of ions in the solution is best determined using the Standard Addition method rather than direct potentiometry.

Apparatus Required

Ion-Selective Electrode for calcium ion (ELIT 8041 – PVC membrane)

Reference electrode: single junction silver chloride (ELIT 001)

Dual electrode head (ELIT 201)

Standard solution: 1000 ppm Ca as CaCl2

ELIT Computer Interface/Ion Analyser, or Ion/pH/mV meter.

100 or 150 ml polypropylene beakers, 100ml volumetric flask, 1, 2, 5, 10ml pipettes.

Calibration

Before use, the electrodes must be calibrated to find an accurate value for the slope under the current operating conditions and in the range of the expected sample concentrations. If sample concentration is completely unknown then it is best to make a quick sample measurement first, using direct potentiometry with an old calibration. Then make up and measure two standards which span the expected sample range - say 1 & 10ppm or 10 & 100ppm - or more closely spaced if working below the linear range of the electrode.

Standard Selection

Enter the slope and expected sample concentration and volume into the Standard Addition software to calculate an appropriate standard concentration and volume. Make up the calculated standard concentration by serial dilution of the 1000ppm standard.

Sample Measurement

1) Pipette the pre-determined sample volume (usually 50 or 100ml) into a beaker, immerse the electrodes, swirl the solution, and record the stable voltage after a few minutes of stabilization.

2) Add the calculated volume of standard, swirl or stirr well then wait for a new stable voltage when the solutions have completely mixed and the electrodes have re-equilibrated.

The software will then calculate the sample concentration and give an estimate of the quality of the determination - if the sample concentration is very different from the expected value then you will be prompted to make a second measurement using a more appropriate volume or concentration of standard.





RETURN TO TOP

Procedure for determining the concentration of CALCIUM (Ca++) in Meat Products

Calcium is extracted from meat directly into distilled water and measured by direct potentiometry.

Apparatus Required

Ion-Selective Electrode for calcium ion (ELIT 8041 – PVC membrane)

Reference electrode: single junction silver chloride (ELIT 001)

Dual electrode head (ELIT 201)

Standard solution: 1000 ppm Ca as CaCl2

Buffer solution (ISAB): 0.1Molar KCL

ELIT Computer Interface/Ion Analyser, or Ion/pH/mV meter.

100 or 150 ml polypropylene beakers, 100ml volumetric flask, 1, 2, 5, 10ml pipettes.

Calibration

Before use, the electrodes must be calibrated by measuring a series of known standard solutions, made by serial dilution of the 1000ppm standard solution. For a full calibration, prepare 100ml of solutions containing 1000, 100, 10, 1, and 0.1ppm Ca. If the approximate range of concentrations of the samples is known then it is only necessary to make two solutions which span this range: e.g. if sample solutions are known to lie between, say, 30 and 130ppm then you could use standards of 10 and 200ppm.

Each standard must be mixed thoroughly with an equal volume of buffer solution in order to compensate for different activity coefficients between samples and standards.

Follow the instructions on the computer interface software to measure these standard solutions and prepare a calibration graph.

Sample Preparation

Weigh out accurately about 1g of meat sample and liquidise with 100ml of distilled water. Centrifuge the mixture and take 25 mls of the supernatant and mix with 25 mls of ISAB.

Sample Measurement

Follow the instructions in the computer interface software and/or electrode operating instructions to measure a series of samples and record the results. Briefly, it is important to note that the electrodes must be washed and dried between each sample, to avoid cross contamination, and sufficient time must be allowed (2 or 3 minutes), before taking a reading after immersion, to permit the electrode signal to reach a stable value. For the highest precision, frequent recalibration is recommended (see operating instructions).

Results

The results will be displayed as ppm and mol/l in the solution. Since the buffer solution has been added equally to standards and samples then these figures will not need adjusting because they will all be affected by the same dilution factor. To obtain the ppm Ca in the sample (mg/Kg), the ppm in the solution must be multiplied by 100 and divided by the sample weight.





RETURN TO TOP

Procedure for determining the concentration of CALCIUM (Ca++) in SKIMMED MILK

Because of the complex matrix, high ionic strength, and other interfering factors in milk, the concentration of ions in the solution is best determined using the Sample Addition method rather than direct potentiometry. This method requires the sample concentration to be known approximately (within a factor of 10) beforehand and involves adding a pre-calculated measured volume of sample to a measured volume of appropriate concentration standard. Nevertheless, if the sample concentration is completely unknown it can be determined with a quick potentiometry measurement first, and the calculations are made relatively easy by the Elit Interface software which determines what volumes and concentrations to use and assesses the quality of the final result so that a repeat measurement can be made with more appropriate volumes if the initial estimate of the sample was badly wrong.

Apparatus Required

Ion-Selective Electrode for calcium ion (ELIT 8041 – PVC membrane)

Reference electrode: single junction silver chloride (ELIT 001)

Dual electrode head (ELIT 201)

Standard solution: 1000 ppm Ca as CaCl2

ELIT Computer Interface/Ion Analyser, or Ion/pH/mV meter.

100 or 150 ml polypropylene beakers, 100ml volumetric flask, 1, 2, 5, 10ml pipettes.

Calibration

Before use, the electrodes must be calibrated to find an accurate value for the slope under the current operating conditions and in the range of the expected sample concentrations. If sample concentration is completely unknown then it is best to make a quick sample measurement first, using direct potentiometry with an old calibration. Then make up and measure two standards which span the expected sample range - say 1 & 10ppm or 10 & 100ppm - or more closely spaced if working below the linear range of the electrode.

Standard Selection

Enter the slope and expected sample concentration and volume into the Sample Addition software to calculate an appropriate standard concentration and volume. Make up the calculated standard concentration by serial dilution of the 1000ppm standard.

Sample Measurement

1) Pipette the pre-determined sample volume (usually 50 or 100ml) into a beaker, immerse the electrodes, swirl the solution, and record the stable voltage after a few minutes of stabilization.

2) Add the calculated volume of standard, swirl or stirr well then wait for a new stable voltage when the solutions have completely mixed and the electrodes have re-equilibrated.

The software will then calculate the sample concentration and give an estimate of the quality of the determination - if the sample concentration is very different from the expected value then you will be prompted to make a second measurement using a more appropriate volume or concentration of standard.





RETURN TO TOP

Procedure for determining the concentration of CALCIUM (Ca++) in SOIL

Calcium is extracted from soil by shaking with sodium acetate solution and measured by direct potentiometry.

Apparatus Required

Ion-Selective Electrode for calcium ion (ELIT 8041 – PVC membrane)

Reference electrode: single junction silver chloride (ELIT 001)

Dual electrode head (ELIT 201)

Standard solution: 1000 ppm Ca as CaCl2

Buffer solution (ISAB): 4 Molar KCl.

Extraction solution: 0.5M CH3COONa (41g/l)

ELIT Computer Interface/Ion Analyser, or Ion/pH/mV meter.

100 or 150 ml polypropylene beakers, 100ml volumetric flask, 1, 2, 5, 10ml pipettes.

Calibration

Before use, the electrodes must be calibrated by measuring a series of known standard solutions, made by serial dilution of the 1000ppm standard solution. For a full calibration, prepare 100ml of solutions containing 1000, 100, 10, 1, and 0.1ppm Ca. If the approximate range of concentrations of the samples is known, and this is within the specified linear range of the ISE, then it is only necessary to make two solutions which span this range: e.g. if samples are known to lie between, say, 30 and 130ppm then you could use standards of 10 and 200ppm or even 20 and 150ppm.

NB: In order to maintain a similar matrix to the sample solutions, each standard must be mixed in the ratio 1:1 with 0.1M sodium acetate (dilute 200mls of the extraction solution to 1 litre with deionised water), Then add 2 ml of KCl buffer solution to 100ml of this mixture, and mix thoroughly before measurement.

Follow the instructions on the computer interface software to measure these standard solutions and prepare a calibration graph.

Sample Preparation

Weigh accurately about 4g of ground, air-dried soil into a screw capped centrifuge bottle and add 50ml of 0.5M sodium acetate extraction solution. Shake mechanically for about 2 hours and then centrifuge to remove all soil particles from the solution. Pipette 10mls of the supernatant into a 100ml flask and dilute to the mark with deionised water. Transfer to a 150ml beaker and add 2ml of KCl buffer and mix well.

Sample Measurement

Follow the instructions in the computer interface software and/or electrode operating instructions to measure a series of samples and record the results. Briefly, it is important to note that the electrodes must be washed and dried between each sample, to avoid cross contamination, and sufficient time must be allowed (2 or 3 minutes), before taking a reading after immersion, to permit the electrode signal to reach a stable value. For the highest precision, frequent recalibration is recommended (see operating instructions). NB: If a stable reading cannot be achieved in a still solution then it is recommended that measurements are taken whilst continually stirring on a magnetic stirrer.

Results

The results will be displayed as ppm and mol/l in the measured solution. Since buffer solution has been added equally to standards and samples these figures will not need adjusting for this dilution factor, but the ppm figure must be multiplied by 250 to compensate for the dilution of the sample and the standard, and divided by the sample weight in order to get the concentration in the soil in mg/Kg.

RETURN TO TOP